AFM Probe Model: Tap190DLC |
| application: |
Tapping Mode,
Intermittent Contact Mode,
Long Cantilever |
 |
| general: |
Rotated Monolithic silicon probe
Symmetric tip shape
Chipsize 3.4 x 1.6 x 0.3 mm
Alignment Grooves |
 |
| coating: |
Diamond-Like-Carbon coating on tip side of the cantilever, 15nm thick;
Aluminum coating on detector side of the cantilever, 30 nm thick |
 |
technical data: |
Resonant Frequency - 190 kHz
Force Constant - 48 N/m |
|
|
value |
range |
| Resonant Frequency |
190 kHz |
± 60 kHz |
| Force Constant |
48 N/m |
20 N/m to 100 N/m |
| Length |
225 µm |
± 12 µm |
| Mean Width |
38 µm |
± 9 µm |
| Thickness |
7 µm |
± 1 µm |
| Tip Height |
17 µm |
± 2 µm |
| Tip Set back |
15 µm |
± 5 µm |
| Tip Radius |
< 15 nm |
| Coating |
Diamond-Like-Carbon coating on tip side of the cantilever, 15nm thick;
Aluminum coating on detector side of the cantilever, 30 nm thick; |
| Half Cone Angle |
20°-25° along cantilever axis 25°-30° from side 10° at the apex |
|
|
|
|